LAURA® Smart
LAURA® Smart is an effective and powerfull solution for the analysis of urine, which is designed for laboratories, ambulances of general practitioners, family doctors or specialists. Evaluation of the urine with LAURA Smart reader helps to eliminate any subjective interpretation of the colour reaction of the diagnostic pad s and therefore remarkably contributes to the correct diagnostic of the patient.
Four types of diagnostic strips, which differ in the number of parameters, are intended for use with the device:
LEU | NIT | pH | ALB | GLU | UBG | BIL | KET | BLD | SG | CF | |
DiaPHAN® LAURA | • | • | • | ||||||||
PentaPHAN® LAURA | • | • | • | • | • | • | |||||
HeptaPHAN® LAURA | • | • | • | • | • | • | • | • | |||
DekaPHAN® LAURA | • | • | • | • | • | • | • | • | • | • | • |
Parameters which can be determined by using diagnostic strips are: pH, protein (ALB), glucose (GLU), ketones (KET), urobilinogen (UBG), bilirubin (BIL), blood (BLD), as well as specific gravity (SG), nitrites (NIT) and leucocytes (LEU). LAURA
Smart identifies the type of the strip used/inserted automatically. The compensation field (CF) allows eliminate the different colours of the urine samples.LAURA Smart has two possibilities of incubation:
Inside of the Reader- Capacity of measurement 60 strips/hour
Outside of the Reader – Capacity of measurement till to 240 strips/hour (special function SmartTiming )
The goal of Smart Timing
The SW enables (turns to green) a new timer in every 15 sec; 4 strips could be measured in 1 minute (240/hour)
LAURA
Smart automatically check itself after each switch-on. The calibration of the reader LAURA is checking by the control mode from SW menu. The special control strips are using for the checkings of the calibration and the right function of the optical parts. The control grey strips are a part of a delivery or it is possible to buy them separately also.LAURA
Smart has double possibilities of power – portable mode (6x 1.5V batteries AA size) and stationary mode (electric power). Other advantages of this reader are touch screen, possibility of connection to the PC or to the laboratory information system (LlS), possibility of connection to the bar code reader (BCR) or to the external keyboard. Reader has the wide choice software menu with a lot of settings according to the user. The user has possibility to choose between different ways of entering the patients’ samples, choose units by the every analysed parameter and add short commentary to the each result. The results are possible to print on a built-in thermo printer with speed 2s/parameter.Technical Data
Description of the device:
- Semi-quantitive reflex photometer
- Measuring using LED + colour detector
- Wavelengths: 470, 540, 650 nm
- Latency: 55 s
- Capacity: 60 strips per hour or 240 strips per hour
- Touch screen (LCD display): 320×240 colour TFT
- Built-in thermo printer, 2sec / parameter
- External communication to PC / LIS via RS 232 or USB port
- Possible connection to external keyboard or a bar code reader
- Memory capacity: last 200 measurements
- Languages for end-users: EN, DE, RU, PL, CZ, HU, IT, FR
PHAN
strips intended for analysis:- DekaPHAN LAURA
- HeptaPHAN LAURA
- PentaPHAN LAURA
- DiaPHAN LAURA
Working Conditions
- External power source 7.5V DC switch adapter 100-240 V / 50-60 Hz
- Internal power source – battery 6x 1,5V AA size
- Power consumption: 20W max / 1 W standby
- Temperature: optimum range 15-35 °C, relative humidity 20-80 %
- Weight: 0,8 kg without adapter, battery and paper roll
- Dimensions: 230 x 127 x 110 mm